10
Installation Drawings
Mounting bracket holes
4 - Ø .18
2.76
2.36
2.76
3.15
2.3
1.06
.47
4.1
psi
6. DIMENSIONS
7. WIRING
Power supply requirements, 12 -32 Vdc, note installation
recommendations below.
7.1 Cable / Wire Specifications
7.2 Wiring Instructions
To reduce potential for noise do not run pressure transmitter cable /
wires alongside (same conduit as) high voltage (line power) lines. For
optimum results use dedicated conduit for GC51 cable / wires.
If using the Cable Gland termination option must use cable within pre-
viously noted diameters to maintain environmental ratings.
Terminal Strip
Cable Requirements
SMKDSP1.5/2-5.08 Phoenix contac
Two core shielded cable
Cable outer diameter: 0.35 to 0.47˝
(9-12mm). Required for proper installa-
tion with cable gland option
Wire Gauge: 14-22 AWG (multi-strand
or solid)