Basic Key Operations
For setting up in each setting mode, the contents of selection UP/DOWN keys are chosen, respec-
t
ively. In all setting modes, values are set with the UP/DOWN keys. Use UP key to increase and
DOWN key to decrease values. A repeat state occurs in three phases of speed when the UP or
DOWN keys are pressed for more than 0.5seconds to increase or decrease numerical value.
Adjusting Zero Point
In Measurement Mode, press the MODE and DOWN keys for greater than 3 seconds; this will dis-
p
lay “_ _ _ after releasing pressure from the pressure port. Approximately 1 second later an auto-
matic zero adjustment takes place and the pressure displays as zero.
When the zero adjustment is successful
Adj appears on the LED display.
The error
E-0 displays for one second when applied pressure was outside the range of -10
10% F.S. , zero adjustment does not happen.
Peak/Bottom (Max./Min.) Hold Display Mode
The GC35 unit keeps the maximum and minimum pressure level applied to the pressure port as
peak and bottom values respectively, in the internal memory. The peak and bottom values are dis-
played as long as press and hold the
or keys respectively. When you select this operation,
the message
PE+ is displayed for one second and selected Peak/Bottom value is displayed.
Peak and bottom values are reset when you reset power to the unit, or by following procedure.
Resetting peak value: While holding the
key, press the key.
Resetting bottom value: While holding the
key, press the key.
Key Lock
Key operations can be nullified to prevent inadvertent overwriting of setting values. Once the key
lock state is set, the mode cannot be shifted to the one other than the peak hold display mode.
The key lock mode cannot be reset by restoring power. It is reset by the following release opera-
tion of a key lock.
Setting of key lock:In measurement mode press
MODE and keys together.
Lo( The lock message is displayed for one second indicating that the unit is locked.
Release of key lock: Press
MODE and keys together.
UnL The Unlock message is displayed for one second indicating that the unit is unlocked
Backup of Setting Values
The unit has an internal EEPROM and settings and the key lock state are maintained even after
the power is turned OFF. Peak / bottom (max./min.) values will not be maintained.
Maximum/Minimum Pressure Capture
The GC35 unit keeps the maximum and minimum pressure level applied to the pressure port
as peak and bottom values respectively, in the internal memory. The peak and bottom values
are displayed while holding the
or keys respectively. Message
PE+ is displayed for
one second and selected Max/Min value is displayed by this operation. Maximum and mini-
mum values are reset when you reset power to the unit or by the following procedure:
Resetting Maximum value: While holding the
key, press the key.
Resetting Minimum value: While holding the
key, press the key.
13. ERROR DISPLAY
An error message and pressure value are alternately displayed when one of the following
errors occurs in the measurement mode (In case of “Out of pressure display range error, only
error message is displayed). Check the content of error message and take the action below
immediately.